Difference between revisions of "User Manual - NIM Spectroscopy Amplifer and Single Channel Analyser"
Jonpo@uio.no (talk | contribs) |
|||
(4 intermediate revisions by 2 users not shown) | |||
Line 1: | Line 1: | ||
+ | [[Image:Frontpanel.jpg|right|Dront panel of an AMP/TSCA module from Canberra manual.]] | ||
+ | |||
This manual was written as part of a student lab exercise in the KJM-FYS 5920 course at the University of Oslo in the autumn 2010. | This manual was written as part of a student lab exercise in the KJM-FYS 5920 course at the University of Oslo in the autumn 2010. | ||
− | Main author (student): | + | Quick links: [[KJM-FYS 5920 Lab Exercise 4 - Student Written Equipment User Manuals|KJM-FYS 5920 Lab Exercise 4]] | [[KJM-FYS 5920 @ UiO - Autumn 2010|KJM-FYS 5920 Course]] |
+ | |||
+ | Main author (student): Pejman Mansouri | ||
Co-author (student): | Co-author (student): | ||
Line 9: | Line 13: | ||
Responsible teacher: Prof. [mailto:j.p.omtvedt@kjemi.uio.no Jon Petter Omtvedt] | Responsible teacher: Prof. [mailto:j.p.omtvedt@kjemi.uio.no Jon Petter Omtvedt] | ||
− | [[Category: | + | I this experiment a CANBERRA AMP/TSCA model 2015A was used. |
+ | |||
+ | The AMP/TSCA or signal amplifier, time single channel analyser is a multi-purpose NIM module. The input signal is a unipolar signal from a pre-amplifier The signal from the pre-amplifier varies in height. The polarity of the input signal should be set by a switch marked as “Input” and +/- on it. | ||
+ | |||
+ | The SCA produces a logical signal whenever the input signal is within a user defined energy window. The energy window is set by two dials in front panel. One is “lower level energy” and the other one is delta-E. Any input signal within this window will generate a logical signal. The logical signal is used for gating. | ||
+ | |||
+ | The logical signals amplitude is independent from the input signals, and is always 5 V. The signal is 500 ns long and is delayed about 200 ns from the top of the amplifier signal. A Gate and Delay module is used to process this signal further before use in a multi channel analyser | ||
+ | |||
+ | The other function of the AMP/TSCA is as an amplifier. The input signal is then <br>amplified by a factor of up to 1280 times. The amplification is adjusted by two knobs on the front panel: “Gain” and “Fine Gain”. The gain sets the amplification by a fixed factor. The fine gain gives a fine, continuous, adjustment of the amplification in addition to Gain. | ||
+ | |||
+ | This output signal is bipolar signal and is a “slow” signal. While the input signal from a pre-amplifier has a typical rise time of 10-20 ns, the output signal has a rise time of 1-2 µs. <br>The fall time of a bipolar signal is measured from the tail part (the negative part) of it with the same criteria as a mono polar signal. | ||
+ | |||
+ | On the front panel is a pole zero adjuster marked as “P/Z” and is operated by a screwdriver. The pole zero is seen on the output signal as a hump on the tail of the signal. | ||
+ | |||
+ | The output signal from the amplifier is sent to a amplifier and delay module.<br> | ||
+ | |||
+ | [[Category:User_manual]][[Category:Electronics]] |
Latest revision as of 12:25, 28 June 2012
This manual was written as part of a student lab exercise in the KJM-FYS 5920 course at the University of Oslo in the autumn 2010.
Quick links: KJM-FYS 5920 Lab Exercise 4 | KJM-FYS 5920 Course
Main author (student): Pejman Mansouri
Co-author (student):
Lab supervisor: Hilde-Therese Nyhus (PhD-student)
Responsible teacher: Prof. Jon Petter Omtvedt
I this experiment a CANBERRA AMP/TSCA model 2015A was used.
The AMP/TSCA or signal amplifier, time single channel analyser is a multi-purpose NIM module. The input signal is a unipolar signal from a pre-amplifier The signal from the pre-amplifier varies in height. The polarity of the input signal should be set by a switch marked as “Input” and +/- on it.
The SCA produces a logical signal whenever the input signal is within a user defined energy window. The energy window is set by two dials in front panel. One is “lower level energy” and the other one is delta-E. Any input signal within this window will generate a logical signal. The logical signal is used for gating.
The logical signals amplitude is independent from the input signals, and is always 5 V. The signal is 500 ns long and is delayed about 200 ns from the top of the amplifier signal. A Gate and Delay module is used to process this signal further before use in a multi channel analyser
The other function of the AMP/TSCA is as an amplifier. The input signal is then
amplified by a factor of up to 1280 times. The amplification is adjusted by two knobs on the front panel: “Gain” and “Fine Gain”. The gain sets the amplification by a fixed factor. The fine gain gives a fine, continuous, adjustment of the amplification in addition to Gain.
This output signal is bipolar signal and is a “slow” signal. While the input signal from a pre-amplifier has a typical rise time of 10-20 ns, the output signal has a rise time of 1-2 µs.
The fall time of a bipolar signal is measured from the tail part (the negative part) of it with the same criteria as a mono polar signal.
On the front panel is a pole zero adjuster marked as “P/Z” and is operated by a screwdriver. The pole zero is seen on the output signal as a hump on the tail of the signal.
The output signal from the amplifier is sent to a amplifier and delay module.