This exercise will teach you to use a NaI-detector to take a spectrum of the <math>\gamma</math> from <sup>99m</sup>Tc.  You learn to analyse the sectrum to determine the amount of <sup>99m</sup>Tc in the sample.
 
This exercise will teach you to use a NaI-detector to take a spectrum of the <math>\gamma</math> from <sup>99m</sup>Tc.  You learn to analyse the sectrum to determine the amount of <sup>99m</sup>Tc in the sample.

Revision as of 08:44, 28 June 2012

Return to main

This exercise will teach you to use a NaI-detector to take a spectrum of the [math]\gamma[/math] from 99mTc. You learn to analyse the sectrum to determine the amount of 99mTc in the sample.

Remember gloves, glasses and lab-coat. If there is some spillage on the gloves control them and immediately change them . Contaminated gloves goes in the designated waste bucket for solid active waste.

Extract 0.5 mL of 99mTc from the solution you received from the counsellor with a automatic pipette and transfer to a 20 mL glass vial put on the lid. Throw away the pipette tip in the designated waste bucket for solid radioactive waste.

Control the outside of the glass vial by drying it with some paper then measure the paper for contamination with the hand held detector

when you are certain that the glass vial is not contaminated on the outside you can take it to the counting room. The counsellor will show you how you use the NaI detector and the [math]\gamma[/math] spectroscopy analytical program Maestro.

Set the counter to count in 180 sec to measure the background ( don't put your glass vial infront of the detector) print out the background spectrum.

Delete the background spectrum and put your sample of 99mTc in front of the counter. Take a new spectrum of your sample set it to count for 180 sec when it is done print out this spectrum. The Tc top is marked with red if it not marked the tops should be around 322 keV. Press the left mouse button on the top and you will receive data on it. Number of counts, N, in the top (“Net Area”): ….............................
total number of counts, Ntot, in the marked area(“gross area”):.................
live time : …........................sec
count rate, R=N/t: …..................sec
energy of [math]\gamma[/math] -gamma top ................ keV
will the background give false counts in the 99mTc-top? (ask counsellor) Explain why/why not with a sketch: