(Created page with "Red Numbers: Present Temperature, Green Numbers: Set point temperature Buttons: P, Arrows: UP and Down * P: change settings and confirm, * UP and Down: change numbers. Alw...")
 
 
(2 intermediate revisions by the same user not shown)
Line 10: Line 10:
  
 
P --->Temperature: Pr.51 65->P --->Time: Pr.t1 15->P --->Temperature: Pr.52 95 ->P --->Time: Pr.t2 99.99->P ---->Temperature: Pr.53 95->P ---->Time: Pr.t2 20->P
 
P --->Temperature: Pr.51 65->P --->Time: Pr.t1 15->P --->Temperature: Pr.52 95 ->P --->Time: Pr.t2 99.99->P ---->Temperature: Pr.53 95->P ---->Time: Pr.t2 20->P
 +
 +
'''NBNB!! Needs update: Initial temperature: set to room temp. Does it go to this temp afterwards? If not, do this explicitly!'''
  
 
Note: The time is correct with actual timer check.
 
Note: The time is correct with actual timer check.
Line 16: Line 18:
  
 
Press and hold U till the greed dot shows, then release. (behind the green number)
 
Press and hold U till the greed dot shows, then release. (behind the green number)
 +
 +
The programmable hotplate is not reliable. '''Please check'''
 +
*    Two times in a row the same procedure for 25 um thick substrate with VWR and programmable hot plates gave good results for VWR hotplate but overdevelopment for programmable hotplate.
 +
*    I have observed that the programmable hotplate seems to spend too long time at each step.

Latest revision as of 15:59, 15 November 2023

Red Numbers: Present Temperature,

Green Numbers: Set point temperature

Buttons: P, Arrows: UP and Down

  • P: change settings and confirm,
  • UP and Down: change numbers.

Always press P to change and confirm settings.

P --->Temperature: Pr.51 65->P --->Time: Pr.t1 15->P --->Temperature: Pr.52 95 ->P --->Time: Pr.t2 99.99->P ---->Temperature: Pr.53 95->P ---->Time: Pr.t2 20->P

NBNB!! Needs update: Initial temperature: set to room temp. Does it go to this temp afterwards? If not, do this explicitly!

Note: The time is correct with actual timer check.

Start the program:

Press and hold U till the greed dot shows, then release. (behind the green number)

The programmable hotplate is not reliable. Please check

  • Two times in a row the same procedure for 25 um thick substrate with VWR and programmable hot plates gave good results for VWR hotplate but overdevelopment for programmable hotplate.
  • I have observed that the programmable hotplate seems to spend too long time at each step.