GM1060, 5um,  
+
GM1060, 5um, 4120 rpm-40sec
  
 
wafer 1: plasma treatment >8 mins
 
wafer 1: plasma treatment >8 mins
  
 
wafer 2: 135 C 20 mins
 
wafer 2: 135 C 20 mins
 +
 +
(Spin coater setting was changed, there was no program 1- slow speed 400 rpm for 20 sec)
  
 
65-5, 95-30
 
65-5, 95-30

Revision as of 17:54, 19 June 2024

GM1060, 5um, 4120 rpm-40sec

wafer 1: plasma treatment >8 mins

wafer 2: 135 C 20 mins

(Spin coater setting was changed, there was no program 1- slow speed 400 rpm for 20 sec)

65-5, 95-30

UV-kub-2: 25%, 4 cycles of 4sec

65-5, 95-20

Fresh PGMEA, shake, 1 min

Submerge in IPA, rinse

Result: small features are well observed

Chemical safety precaution: keep all PGMEA-contacted waste inside hood, The hood sash should be raised higher than the thick black line with arrow.

Most frequent chance of spill: 1-during washing of PGMEA, left hand (holding the wafer) will be in contact.

2-when pouring the waste to bottles, PGMEA is spilled outside of the bottle.

Maybe need a bigger funnel? Or a glass container with a spout?