* Started LAF-bench
+
*   Started LAF-bench
* N2 valve had been left open, no pressure left!
+
*   N2 valve had been left open, almost no pressure left!
 +
*    cleaned wafers with plasma only
 +
*    1st wafer
 +
**        3700 RPM ~ 20 um
 +
**        prebake on programmable hot plate
 +
**        Masking with UV-Kub 2. UV at 11% intensity for 30 sec.
 +
**        Hot plate too hot, 65C postbake happened at 72-66C
 +
**        The hot plate seems to take longer than the timer on my telephone. This needs to be checked!
 +
**        development with PGMEA for 2:45 was too long time! Misjudged the residue and continued another 30s, then it was completely destroyed.
 +
*    2nd wafer 3100 RPM ~ 25 um
 +
**        prebake on VWR hot plate, no nice control of ramp in temp
 +
**        Masking with UV-Kub 2. UV at 11% intensity for 30 sec.
 +
**        Postbake on VWR hotplate
 +
**        Developed 3min in PGMEA gave perfect sylinders and gaps between them. quick dip to remove a little residue and done.

Latest revision as of 15:43, 3 May 2021

  • Started LAF-bench
  • N2 valve had been left open, almost no pressure left!
  • cleaned wafers with plasma only
  • 1st wafer
    • 3700 RPM ~ 20 um
    • prebake on programmable hot plate
    • Masking with UV-Kub 2. UV at 11% intensity for 30 sec.
    • Hot plate too hot, 65C postbake happened at 72-66C
    • The hot plate seems to take longer than the timer on my telephone. This needs to be checked!
    • development with PGMEA for 2:45 was too long time! Misjudged the residue and continued another 30s, then it was completely destroyed.
  • 2nd wafer 3100 RPM ~ 25 um
    • prebake on VWR hot plate, no nice control of ramp in temp
    • Masking with UV-Kub 2. UV at 11% intensity for 30 sec.
    • Postbake on VWR hotplate
    • Developed 3min in PGMEA gave perfect sylinders and gaps between them. quick dip to remove a little residue and done.