Difference between revisions of "20211106 microfluidics"

From mn.fysikk.laglivlab
Jump to: navigation, search
Line 8: Line 8:
 
=== Adding new PDMS to wafer ===
 
=== Adding new PDMS to wafer ===
 
- Used 43.70g PDMS and 3.68 g curing agent.
 
- Used 43.70g PDMS and 3.68 g curing agent.
 +
 +
 +
=== Troubles with microfluidic chips ===
 +
We had a few problem with punching the chips: starting with Anniken's (?), after some time after plasma it seemed that the chip closed itself (the part facing the air, approx half way in the ). We then tried to re-punch holes out *from the top* which seemed to work.
 +
To be sure, we cut out another chip (Thomas') but after plasma it turned out to be the same problem. We still managed to get tubes in, watertight.
 +
 +
==== Why? ====
 +
- these PDMS chip were coming from the same wafer (broken BTW), which was covered which a very thick layer of PDMS

Revision as of 16:13, 11 June 2021

Preparing the chip & plasma cleaning

- Cut out Vilde and Dag's chip designs.

- Made 1.6mm holes in the inlets&outlets

- Plasma cleaned

Adding new PDMS to wafer

- Used 43.70g PDMS and 3.68 g curing agent.


Troubles with microfluidic chips

We had a few problem with punching the chips: starting with Anniken's (?), after some time after plasma it seemed that the chip closed itself (the part facing the air, approx half way in the ). We then tried to re-punch holes out *from the top* which seemed to work. To be sure, we cut out another chip (Thomas') but after plasma it turned out to be the same problem. We still managed to get tubes in, watertight.

Why?

- these PDMS chip were coming from the same wafer (broken BTW), which was covered which a very thick layer of PDMS