Difference between revisions of "20210309"
From mn.fysikk.laglivlab
Dagkd@uio.no (talk | contribs) m |
Dagkd@uio.no (talk | contribs) (contiuous filling in) |
||
Line 14: | Line 14: | ||
* Use new wafer and go straight for plasma cleaning | * Use new wafer and go straight for plasma cleaning | ||
* Skip heating and do spinning | * Skip heating and do spinning | ||
+ | |||
+ | I lost all my edits! Obviously it's important to save before leaving idle | ||
+ | |||
+ | * SU8 to wafer using syringe worked perfectly. | ||
+ | * Used a single spin: ramp 100, 900 rpm at 40s | ||
+ | * Display said it only spun at 830 RPM | ||
+ | * Noticeable bulge at edge of wafer. Tried to wait for an hour, but it did not change much. | ||
+ | * Transferred to hot plate: 65C 15min, 95C 90 min, should be 120, but display only has 2 digits. time manually. |
Revision as of 12:21, 9 March 2021
Continued from yesterday
- Had left the insolated wafer on the hot plate over night. This means 15 hours rest instead of 10 min.
- Table says 5 min develop.
- I used new PGMEA
- After 3 minutes I noticed white structures in the region around the main structures. I did not understand what this signified
- After 5 min I rinsed with isopropanol (IP)
- I struggled to get good view in the microscope. Therefore I miscalculated what I saw and thought I needed to develop more.
- I developed about 30 s more.
- In the microscope it was now clear that the structures where overdeveloped and all the smallest features where erased.
Time to start again
New try
- Use new wafer and go straight for plasma cleaning
- Skip heating and do spinning
I lost all my edits! Obviously it's important to save before leaving idle
- SU8 to wafer using syringe worked perfectly.
- Used a single spin: ramp 100, 900 rpm at 40s
- Display said it only spun at 830 RPM
- Noticeable bulge at edge of wafer. Tried to wait for an hour, but it did not change much.
- Transferred to hot plate: 65C 15min, 95C 90 min, should be 120, but display only has 2 digits. time manually.