Present: Lasse, Morten, Elizabeth, Catherine
 
Present: Lasse, Morten, Elizabeth, Catherine
  
 +
==== Results (rotations per 10sec): ====
 +
Each row for a clip is a different cell.
 
{| class="wikitable"
 
{| class="wikitable"
 
!10kHz
 
!10kHz
Line 15: Line 17:
 
!50MHz
 
!50MHz
 
|-
 
|-
|0
+
|0.25
|0
+
|3
|0
+
|3.5
|0
+
|4.25
|0
+
|3.5
|0
+
|2.75
|0
+
|0.25
|0
+
|0.25
|0
+
| -0.375
|0
+
| -0.625
|0
+
| -1
|0
+
| -1.25
 
|-
 
|-
|0
+
|0.125
|0
+
|2.25
|0
+
|2.25
|0
+
|3.25
|0
+
|3
|0
+
|2
|0
+
|1.125
|0
+
|0.375
|0
+
| -0.25
|0
+
| -0.625
|0
+
| -1
|0
+
| -1
 
|-
 
|-
|0
+
|0.125
|0
+
|2
|0
+
|3
|0
+
|3
|0
+
|3
|0
+
|3.5
|0
+
|0.625
|0
+
|0.25
|0
+
| -0.25
|0
+
| -0.5
|0
+
| -0.75
|0
+
| -0.75
 
|}
 
|}
  
Line 58: Line 60:
  
 
- The rotation chamber surface is really messy, cell remains after 4-5 wash cycles with isopropanol and ultrasound bath. We should consider improving cleaning protocol, with for instance compressed air/nitrogen or something.
 
- The rotation chamber surface is really messy, cell remains after 4-5 wash cycles with isopropanol and ultrasound bath. We should consider improving cleaning protocol, with for instance compressed air/nitrogen or something.
 +
 +
- Agrees fairly well in peak amplitude and peak frequency and turning point as seen in the [[10.03.23]] experiment.
 +
 +
[[File:BV2 Rotation 2.png|left|thumb|720x720px|The data compared to the previous BV2 WT experiment.]]

Latest revision as of 13:20, 21 March 2023

Present: Lasse, Morten, Elizabeth, Catherine

Results (rotations per 10sec):

Each row for a clip is a different cell.

10kHz 100kHz 200kHz 300kHz 500kHz 1MHz 5Mhz 10MHz 20MHz 30MHz 40MHz 50MHz
0.25 3 3.5 4.25 3.5 2.75 0.25 0.25 -0.375 -0.625 -1 -1.25
0.125 2.25 2.25 3.25 3 2 1.125 0.375 -0.25 -0.625 -1 -1
0.125 2 3 3 3 3.5 0.625 0.25 -0.25 -0.5 -0.75 -0.75

Observations:

- The rotation chamber surface is really messy, cell remains after 4-5 wash cycles with isopropanol and ultrasound bath. We should consider improving cleaning protocol, with for instance compressed air/nitrogen or something.

- Agrees fairly well in peak amplitude and peak frequency and turning point as seen in the 10.03.23 experiment.

The data compared to the previous BV2 WT experiment.