Started 08:30 following procedure on March 8th.

Participants present: Dag, Adam, Kristina and one more person from the department.

  • New box of 3" wafers, cleaned 2 wafers with IP and N2,
  • plasma cleaning ~3 min (100 on timer setting) 100%
  • on VWR hot plate 09:30 set to 150C, 10 min
  • SU8 from syringe (40 ml left)
  • Spread at 400, 40s, set 1760, reading 1690 40s -> 50 um
  • 2 small dust specs on second wafer, 1st wafer seemed fine
  • Put on to programmable hot plate
  • Hot plate started heating too early, relaxation after spinning only 5-10 min
  • 15 min @ 65 C, 120 min @ 95 C, started 10:30
  • The surfaces are not completely flat. Probably due to too little relaxation!
  • Exposed 45 s @ 20% of full intensity
  • Returned to hot plate for 15 min @ 65 C, 40 min @ 95 C, started 13:35